1. The heat dissipation problem of the scanning head Scanning head in the work will produce a lot of heat, the scanning head must be installed in the aluminum or copper, such as good heat dissipation bracket, and to ensure that the normal work, scanning head surface temperature can not exceed 45 ℃.
The XY bracket preferably adopts an integrated structure, which can enhance the heat dissipation and improve the stability of the system.
2. Insulation problem The scanning head and the bracket must carry on the insulation treatment, in order to prevent the static electricity, the high pressure to the scan head damage, enhances the system to resist the disturbance the ability, but at the same time cannot affect the scanning head the heat dissipation. Therefore, the company in the Annex bag provides insulation heat-conductive film, please be sure to use the user.
3. When designing an XY bracket, it is important to note that lenses cannot be touched when the XY scan head is at the same time as the maximum angle, otherwise the lenses and scan heads will be damaged.